FAQs 32. What safety features are incorporated with the UV light in the SterilGARD(r) II & SterilGARD(r) III? So many of our customers request UV lighting that we've made it standard on the SterilGARD(r)II & SterilGARD(r)II. We've built in safety features to protect cabinet users and ensure that the light is used properly. The UV/germicidal light switch and cabinet lighting cannot be turned on simultaneously. A UV interlock switch assures that UV illumination occurs only when the sash is fully closed. To operate the UV light, leave the blower on, lower the sash until it's fully closed, turn off the fluorescent lights and then turn on the UV light. - Updated: July 31, 1999 Have