24. Do I have to duct my SterilGARD exhaust outdoors?

The SterilGARD does not have to be connected to an exhaust duct to provide its rated level of performance; however, you or your safety officer may want to confirm that the installation meets all relevant local and state codes pertaining to your application. Obviously, it's important to consider carefully the agents you'll be working with to ensure the appropriate level of product, personnel and environmental protection. Again, your safety officer can be most helpful in meeting these concerns. If external ducting is indicated, your SterilGARD can be readily adapted. When not ducted, SterilGARD is classified as Class II Type A1. When ducted, it becomes a Class II Type A2. - Updated: July 26, 2002