SterilchemGARD III Advance° Class II, Type B2 Biological Safety Cabinet


SterilchemGARD®III Advance°
Class II, Type B2, 100% Exhaust Vertical Flow Biological Safety Cabinet
The SterilchemGARD® III Advance° Class II, Type B2 total exhaust cabinet produces a HEPA-filtered, particulate-free airflow throughout the work area. Typical applications include toxicology laboratories and similar facilities where microbiology, cell culture and pharmaceutical procedures are common. Model SterilchemGARD® III Advance° SG403ATX shown.

CSA
Listed
Meets
NSF
Standards
Underwriter
Laboratories
Listed
UniPressure™
Preflow
Plenum

Click here to download product literature

SterilchemGARD III Advance° offers 100% total exhaust with integrated HEPA supply and exhaust filters to set the industy’s highest standards for comfort, performance and safety.

In the laboratory environment, safety is everything. Now, with the introduction of SterilchemGARD® III Advance°, The Baker Company offers a new adaptive ergonomic design, combined with a unique airflow management system and proven containment technology to improve comfort, increase productivity and reduce the cost of ownership.

SterilchemGARD® III Advance° is a collective design, representing an esteemed reputation for performance, reliability and craftsmanship associated with The Baker Company worldwide for over 50 years.

SterilchemGARD® III Advance° starts with a simple idea: a better angle. From your first experience with the 10° slanted viewscreen, you’ll appreciate how easy this cabinet is to use. Your work is easy to reach. There’s more functional room inside. Eye-level controls are easy to see. There is less glare. As you work, your head and neck, arms and elbows remain in the most natural position, less tiring than before.

Concealed behind the attractive front panels, you’ll find another Baker exclusive, the UniPressure™ Preflow Plenum, an elegantly simple yet effective air management system that puts more uniform air pressure on the supply filter for better vertical flow across the work surface. A simple idea, to be sure, but by moving the blower/motor to a horizontal orientation from vertical, we’ve created a more efficient, less demanding and quieter airflow system.

And years later, if it comes time for maintenance, your certifier will find the cabinet easy to seal and easy to decontaminate. Our new direct-to-plenum filter mounting technique puts the HEPA filter within easy reach, easy to change.


Advantages Ergonomic Features and Benefits
Comfortable and Easy To Work In • The slanted viewscreen is sloped 10° to create the
most natural head and elbow positions when working, minimizing arm extension required to reach the worksurface and reducing glare.
• The eye-level control panel is positioned to face down toward the user for greater visibility and easier access.
• Reduced front grille depth and a slim profile lower plenum combine to move the work surface closer to the front and your lap for better arm position.
• Our new UniPressure™ Preflow Plenum delivers
quieter, more efficient performance.
• Petcocks are offset, not one on top of the other, for easier access and use.
• An armrest across the cabinet front improves support and comfort.
• A telescoping stand allows the work surface height to be adjusted to 30" or 385/8", nominal.
More Space For Your Work, Your Equipment and You • SterilchemGARD® III Advance° offers the largest
unobstructed, useable work area in the industry.
• Depth of the front grille is minimized to bring the work closer to the user.
• A straight back wall maximizes available work area and accommodates lab equipment and apparatus
commonly used in the biological safety cabinet.
Easy To Clean and Maintain

• A unitized, unobstructed drain pan with radius corners facilitates cleaning.
• Streamlined design of the sliding viewscreen simplifies sealing prior to total cabinet decontamination.
• The UniPressure™ Preflow Plenum simplifies filter changing.
• All electrical control components are consolidated on a panel behind the hinged canopy for singular access outside the containment area.
• The innovative horizontal blower/motor assembly is easily removed and replaced.

 



Exclusive 10° slanted viewscreen– less glare, work easier to reach.

SterilchemGARD® III Advance° is the most exciting development in biological safety cabinets in years. Consider the angle. Learn more about SterilchemGARD® III Advance. It’s what you would expect from The Baker Company.

Design and Performance Features
SterilchemGARD® III Advance incorporates UniPressure™ Preflow, the Baker exclusive high performance airflow system which extends filter life and reduces cost of ownership. This high performance airflow system provides optimum protection from particulates, 100% exhaust, with no air recirculation to the work area.

  • The unique UniPressure™ Preflow Plenum design apportions and distributes air across, then through the HEPA supply filter, improving downflow uniformity, extending useful filter life and reducing noise.
  • The UniPressure™ Preflow filter housing is surrounded by negative pressure to improve containment; any possible gasket leaks are contained under negative pressure.
  • An innovative telescoping filter mount provides a direct seal of the filter to the UniPressure™ Preflow Plenum, and simplifies filter replacement.
  • Front accessible supply and exhaust HEPA filters are rated at 99.99% minimum efficiency in capturing 0.3 micrometer particulates.
  • The Baker exclusive momentum air curtain contains particulates within the work area and inhibits migration of room air into the work area.
  • High velocity return air slots located along the side walls of the work area and along the top of the viewscreen maximize the biological safety cabinet protective capabilities by preventing the escape of particulates, ensuring that no unfiltered air enters the work area, and preventing gases, vapors or particulates from migrating up behind the viewscreen and escaping into the operator’s environment.
  • An aerodynamically designed airfoil at the front opening facilitates the flow of room air into the front grille, ensuring that no room air flows into the work area.
  • A microprocessor controller with touchpad membrane control panel simplifies operation; controls are more reliable, easy to use and easy to clean.
  • An audible/visual sash alarm system warns of unsafe viewscreen opening.
  • The airflow monitor warns of insufficient exhaust airflow.
  • Detachable side panels offer greater flexibility for plumbing considerations; the reinforced panel design reduces the number of fasteners and simplifies removal and replacement.
  • The front panel is lighter and easier to remove for filter changing.


SterilchemGARD® III Advance° Design Features

1. Slanted viewscreen sloped 10° for the most comfortable head and elbow position when working; minimizes arm extension required to reach the work surface.

2. Low profile, unitized drain pan beneath the work surface allows for best arm position above the user’s lap.

3. Eye-level controls face down toward the user for greater visibility and easier access.

4. Microprocessor controller with touchpad membrane control panel simplifies operation.

5. Stand with telescoping legs allows work surface to be adjusted 30" or 385/8", nominal.

6. Audible/visual sash alarm with timer and mute switch encourages safe operation.

7. Airflow monitor warns of insufficient exhaust airflow.

8. High velocity return air slots maximize protective capabilities.

9. Easy access HEPA supply and exhaust filters, 99.99% minimum efficiency for 0.3 micrometer particulates.

10. StediVOLT® voltage regulator automatically adjusts for low and high voltage situations to maintain proper airflow.

11. Energy-saving fluorescent lighting with solid-state ballasts reduces flicker, minimizes heat output, improves eye comfort and extends lamp life by as much as 50%.

12. Non-glare work surface improves comfort.

13. Ground-fault interrupt (GFI) outlets with self-resetting circuit breaker for user safety; trip light warns circuit has been disconnected because of ground fault error.

14. Straight back wall easily accommodates lab apparatus and instrumentation.

Class II, Type B2 Applications
Applications include biological testing and product preparation involving low to moderate risk agents (Biosafety Levels 2 and 3) where chemical effluent is present and clean air is essential.

Because there is no recirculation within the work area, the cabinet may be useful for work with chemicals and gases if required as an adjunct to the microbiological studies that are being conducted.


Total Exhaust, No Recirculation



SterilchemGARD®III Advance° Airflow
Click to enlarge

The SterilchemGARD® III Advance° is ideal for applications which generate small amounts of chemical vapors or gases. The cabinet incorporates a unique negative pressure enclosure which provides additional containment capability.

  • Room air enters through a washable, re-usable prefilter and passes through a blower/motor in the front of the cabinet, then to a HEPA supply filter into the work area as vertical unidirectional airflow.
  • Descending air splits and flows through the perforated front and rear grilles into the space below the worksurface.
  • Simultaneously, room air entering through the front access opening is also pulled through the front perforated grille and all air is immediately exhausted.
  • Contaminated air entering the front and rear grilles is drawn under negative pressure through the stainless steel plenum at the rear of the cabinet, through the exhaust HEPA filter, and out to the building exhaust system.
  • The exhaust removes 100% of air within the work area; all air is pulled into the facility exhaust system for appropriate treatment; there is no recirculation.

SterilchemGARD®III Advance° Containment and Protection
  Type B2 Type B2
  Exhausted to untreated facility exhaust system
Exhausted to treated facility exhaust system
Protection
From
Particulates
Protects personnel, product and the environment Protects personnel, product and the environment
Protection
From Gases
and Vapors
Protects personnel and product Protects personnel, product and the environment

Installed to the facility exhaust system by a hard connection, the SterilchemGARD® III Advance° provides particulate protection, plus additional protection of personnel and product from vapors and gases. If the facility exhaust system is specially treated with activated charcoal, scrubbing or catalytic conversion, protection from exhausted vapors and gases is extended to the environment.


Momentum Air Curtain Increases
Protection Without Restricting User Access



Momentum Air Curtain
Click to enlarge

SterilchemGARD® III Advance° employs a unique momentum air curtain which offers an added measure of containment, personnel and product protection exclusive to the Baker design.

  • This unique airflow design creates a more impenetrable air barrier or “momentum air curtain” at the front of the cabinet, increasing the protective capabilities of the cabinet.
  • The strategic position of the stainless steel diffuser just below the supply filter creates a faster airflow of a nominal 100 FPM at the front of the work zone.
  • Airflow over the center of the work surface is relatively slower at a nominal 55 FPM.
  • The resulting air curtain, high velocity return air slots, aerodynamically contoured front opening surfaces, and optimum air intake velocity combine to minimize turbulence and prevent migration of airborne contaminants into or out of the work area.

High Velocity Return
Air Slots Capture Unfiltered Air

Containment and cleanliness are accomplished with precise control of airflow volumes and velocities. High velocity return air slots, unique in the Baker cabinet design, have been proven to maximize the biological safety cabinet protective capabilities.

  • High velocity return air slots are located along the side walls of the work area.
  • Air is drawn into these slots at 1,000 FPM (nominal), preventing the escape of particulates and ensuring that no unfiltered air enters the work area.
  • High velocity return air slots are located behind the top and on both sides of the viewscreen to prevent gases, vapors or particulates from migrating up behind the viewscreen and escaping into the operator’s environment, and to prevent room air from migrating down behind the viewscreen and contaminating the work area.

Negative Pressure Double-Wall
Plenums Enhance Safety

The all-metal, double-wall design in SterilchemGARD® III Advance° creates base, side and rear wall plenums which capture and contain contaminated air under negative pressure. This unique Baker design feature prevents contaminated air from escaping into the laboratory environment in the event of a puncture or leak in the cabinet.

  • Any plenum which contains contaminated air under positive pressure is completely surrounded by negative pressure areas.
  • Because of negative pressure under the worksurface, side and rear walls, the work area is completely surrounded by negative pressure.


Microprocessor Controls


The microprocessor-based membrane control panel simplifies operation.

The SterilchemGARD® III Advance° microprocessor control system incorporates a unique membrane control panel which is easily cleaned, moisture-proof and exceptionally reliable with no mechanical relays or switches to corrode or fail. A variety of important safety features have been designed into the control system.

  • The control panel is positioned to face down toward the user for greater visibility and easier access.
  • The audible alarm mute button overrides the alarm for 5 minutes when the sash is raised for cleaning, loading or unloading.
  • UV light cannot be activated unless fluorescent light is off and sash is closed.
  • The membrane switch panel is just .03 inch thick, with a smooth surface for cleanability.
  • Large buttons with icons are easy to identify and operate.

 

The SterilchemGARD® III Advance° has a unique negative pressure enclosure which provides additional containment capability. The cabinet work area and plenums containing contaminated air under positive pressure are surrounded by negative pressure areas.


High Visibility Viewscreen With Sash Alarm
SterilchemGARD® III Advance° has a streamlined 10° slanted sliding viewscreen which makes the cabinet easier to use and more comfortable to work in.

  • The operator’s head, neck, arms and elbows remain in the most natural positions to improve comfort and reduce fatigue.
  • All work is within reach; the angled viewscreen design provides a safe, highly visible and easily accessible work area for a range of procedures.
  • The slanted sliding viewscreen minimizes glare.
  • The vertical lift viewscreen is constructed of 1/4" safety plate glass.
  • The sash has an 8" working opening with an average calculated air intake velocity of 100 FPM to 110 FPM.
  • A maximum viewscreen sash opening of 18" simplifies equipment and instrument loading and unloading.
  • An integrated viewscreen sash alarm includes an audible and visual indicator to warn of improper sash opening above the 8" opening.
  • The viewscreen sash alarm timer and audible alarm mute button override the alarm for five minutes when the sash is raised for cleaning, loading or unloading.

UniPressure™ Preflow
Advanced Filter Plenum Design

The Baker exclusive UniPressure™ Preflow Plenum provides quieter, more efficient operation.

UniPressure™ Preflow, a revolutionary air delivery system, offers significant benefits through unique integration of supply and exhaust HEPA filter assemblies, an all-steel plenum and a new horizontal blower/motor configuration.

  • The supply air plenum is configured to provide more uniform airflow to the supply filter, while reducing noise and increasing motor reserve capacity.
  • The all-steel plenum clamps the supply HEPA filter in place. Force is applied to the full perimeter of the filter rather than point force.
  • A closed-cell neoprene gasket forms an airtight seal around the filter periphery.
  • The horizontal blower assembly creates more uniform air direction to the supply HEPA filter.


Supply and Exhaust HEPA Filter Performance
Because HEPA filters remove microorganisms and airborne particulates (generally called aerosols) from the air, the quality, performance and useful life of supply and exhaust filters are critical considerations in the biosafety environment.

  • Each HEPA filter in a Baker cabinet is scan-tested for leaks and tested for overall efficiency by the manufacturer, then individually leak-tested by Baker after installation to assure leak-free performance.
  • Minimum HEPA filter efficiency is 99.99% for particulates of 0.3 micrometers, with increased efficiency for particulates greater and smaller than 0.3 micrometers.
  • Both exhaust and supply HEPA filters can be inserted and removed from the front of the cabinet. This time-saving feature is useful since the SterilchemGARD® III Advance° must be connected to an exhaust system.
  • A closed-cell neoprene gasket provides an airtight seal between the filter assembly and the metal plenum.

Blower/Motor System Helps Extend Filter Life
Baker has selected an optimum blower/motor combination to assure performance and to provide for an average filter life of 7 to 10 years, the longest life in the industry. The Baker split capacitor blower/motor can provide a constant volume of air despite increases in resistance due to filter loading, resulting in extremely long filter life as verified by simulated filter loading tests.

Extended filter life minimizes filter replacement and decontamination costs and diminishes the hazard to personnel by reducing the frequency of opening the contaminated areas of the cabinet. The more filter resistance the blower/motor system can overcome, the longer between filter changes.

  • As particulates fill the filter media, resistance to airflow increases.
  • The Baker blower/motor is a direct drive permanent split capacitor motor designed to deliver the maximum amount of filter life for each cabinet.
  • The blower/motor automatically compensates for filter loading up to 60% above the initial pressure drop across the filter. Beyond that point the blower/motor has a manual speed control which can be turned up to achieve even longer filter life.
  • The filter need not be changed until the blower/motor system cannot deliver adequate air volume to maintain the nominal setpoint, ±5 FPM.

Airflow Monitor,
Automatic Shut-off

  • An audible/visual mass airflow alarm system warns when the exhaust decreases to 20% of the set operating rate.
  • If the low flow condition remains for more than 6 seconds, the alarm will terminate power to the supply blower to prevent pressurization of the work area.
  • If the audible and visual alarm indicates inadequate exhaust flow, all work should be suspended until proper flow is restored.

Easier On-Site Certification and Testing
Several SterilchemGARD® III Advance° design features combine to simplify certification and maintenance for qualified service personnel, thus reducing downtime and improving life cycle costs.

  • The cabinet design puts the supply and exhaust HEPA filters within easy reach.
  • Both exhaust and supply filters can be easily inserted and removed from the front of the cabinet. This time-saving feature is useful since the SterilchemGARD® III Advance° must be connected to an exhaust duct.
  • An optional external, adjustable, airtight damper allows the certifier to compensate for changing resistance of the exhaust filters and provides a positive seal for decontamination; the damper regulates the balance between intake and downflow so proper air circulation ratios are maintained.
  • A manual speed controller allows the certifier to increase blower/motor speed to compensate for supply filter loading.

StediVOLT® Motor Speed Controller
Maintains Optimum Performance

The Baker exclusive StediVOLT® motor speed controller reduces the risk of performance degradation and possible product loss due to line voltage fluctuations.

  • StediVOLT® minimizes changes in cabinet performances due to routine voltage fluctuations, as well as voltage changes caused by high-demand brownout conditions.
  • StediVOLT® “reads” incoming voltage, automatically compensates for low and high voltage and maintains proper blower speed.


Comfortable Lighting Improves Visibility, Reduces Heat
Ergonomic benefits of the SterilchemGARD® III Advance° include“flicker-free” fluorescent lamps and electronic ballasts which provide better visibility, less heat at the face (front) of the cabinet and improved user comfort.

  • Lamps provide more than 80 foot-candle lighting at the work surface.
  • Electronic ballasts eliminate fluorescent lamp “flicker,” minimizing eyestrain and improving productivity.
  • The warm, natural illumination exhibits better color fidelity.
Motor Loading Capacity
Model Pressure Drop Increase Maximum Increase With Manual Blower Speed Control
SG403ATX, 4-foot
SG603ATX, 6-foot
60%
60%
180%
180%
SterilchemGARD® III Advance° is capable of automatically handling a 60% increase in pressure drop across the filter without reducing total air delivery more than 10%. With the use of the manual speed controller, these cabinets can handle up to a 180% increase in pressure drop across the filter.


Optional Ultraviolet Lamp Includes Safety Interlock

An optional ultraviolet germicidal lamp assists in contamination control.

  • The UV light interlock assures that UV illumination occurs only when the sash is fully closed.
  • For added safety, the UV germicidal light switch and cabinet lighting cannot be turned on simultaneously.


Cabinet Construction Enhances
Performance, Minimizes Downtime

Baker cabinet designs represent many years of experience in stainless steel fabrication and craftsmanship. Design considerations such as wide radius corners, aerodynamically shaped surfaces and non-glare satin finish interiors combine to improve comfort, simplify cleaning and maintain proper containment.

  • SterilchemGARD® III Advance° includes a corrosion-resistant, 16-gauge, Type 304 stainless steel work surface with smooth radiused corners to permit easy cleaning. The recessed work surface tray retains spills.
  • The satin finish work surface diminishes harsh light reflection.
  • A protective grille located at the bottom of the rear and side return air plenums prevents wipes and other paper materials from being inadvertently drawn into the blower system. This eliminates costly decontamination and downtime to extract materials from the blower.
  • Work area side walls and rear wall are one-piece, 16-gauge stainless steel construction with 7/16" radiused corners for cleaning convenience.
  • The position of the stainless steel supply air diffuser/filter protector located below the supply filter in the work area creates the momentum air curtain at the front opening and provides filter protection.
  • The 16-gauge cold-rolled steel exterior sealing panels are protected with a white baked enamel finish.
  • The entire cabinet is airtight. Each component is welded, gasketed or assembled with hermetically sealed joints. Each cabinet is bubble tested under pressure to ensure the integrity of the seals.
  • To facilitate cleaning the drain pan, the work surface and supports are easily removed and the drain pan has large 7/16" radii on all sides.

Telescoping Stand
SterilchemGARD® III Advance features a stand with telescoping legs which allows the cabinet to be adjusted to standard work surface heights of 30" or 385/8", nominal.
  Knee Space Height Work Surface Height Overall Height
Minimum Setting 281/8" 301/8" 923/8" - 947/8"
Maximum Setting 365/8" 385/8" 983/8" - 1007/8"


Electrical System Offers Safety and Convenience
The SterilchemGARD® III Advance° electrical system is designed for safety and convenience when using most commonly required laboratory apparatus and instrumentation. A single ground fault interrupter (GFI) protects a pair of duplex outlets mounted on the work area left and right side wall.

  • Each outlet is provided with a drip-proof cover.
  • The internal work area outlets are on a separate circuit from the cabinet lights and blower/motor so that an overload caused by research equipment will not affect cabinet function.
  • Hermetically sealed bulkhead connectors are used to provide reliable containment protection for electrical connectors which penetrate the cabinet contaminated areas.
  • For safety reasons, the SterilchemGARD® III Advance° cabinet utilizes a single power cord and plug arrangement. This assures that a second power source is not left on if maintenance functions are necessary.
  • The SterilchemGARD® III Advance° is UL listed for electrical safety and integrity.


Functional Utilities With Petcocks,
Valves and Plumbing Connections

Plumbing and drainage connections are strategically placed for both convenience and proper air management within the airflow plenums.

  • One petcock is located in the right work area side wall. Three plugged penetrations are provided to accommodate optional petcocks inside the cabinet, one in the right side wall and two in the left.
  • Petcocks and penetrations are offset, not one above the other, for easier access and use.
  • External plumbing connections are made to the bottom (or as a factory option through top or back) of the cabinet rather than through its sides. This arrangement allows installation next to building walls or furniture to the right or left of the cabinet, saving valuable lab space.
  • A stainless steel ball valve provides safe and effective drainage of the drain pan.

Options and Accessories
For convenience, most options, accessories and modifications are factory installed and should be specified when ordering. Commonly requested options are listed below. For detailed information on accessories and modifications, refer to The Baker Company Accessories literature or contact The Baker Company.

  • UV germicidal lamp
  • Stainless steel IV bar
  • Additional petcock (specify label and location)
  • Plumb to top
  • Plumb to rear
  • Plastic storage bins
  • Magnehelic gauge
  • Ergonomic adjustable footrest
  • Airtight damper assembly
  • Lab stool
  • Bench model
  • Seismic restraints
  • Blower/Motor switch (with additional pole for interlocking with house exhaust system)
  • 12” round exhaust collar

To Order
For ordering information, terms and conditions of sale, contact Baker or click here for the name of your authorized Baker representative.

Caution:
A Class II, Type B2, 100% total exhaust biological safety cabinet is suitable for work with low to moderate risk (Biosafety Levels 2 and 3) biological agents. It may also be used with biological agents treated with toxic chemicals and radionuclides required as an adjunct to microbiological studies (as stated by National Sanitation Foundation, International Standard #49).

Note: The adequacy of this containment cabinet should be determined by an industrial hygienist or safety officer.

Site preparation information, architectural drawings, detailed dimensions and purchase specifications are available.


Summary Specification
SterilchemGARD® III Advance°
Class II, Type B2 Biological Safety Cabinet, Vertical Flow
Model Size Opening Electrical Amps/
Breaker
Interior
Dimensions
Exterior
Dimensions
Net Weight Ship Weight
SG403ATX 4' 8"
(18")
115V, AC 9.0/20 46" W
x 237/16" F-B
x 271/4" H
535/8" W
x 309/16" F-B
x 641/4" H
550 lbs 750 lbs
SG603ATX 6' 8"
(18")
115V, AC 12.4/20 70" W
x 237/16" F-B
x 271/4" H
775/8" W
x 309/16" F-B
x 641/4" H
800 lbs 1000 lbs





Amperage Chart
Model Blower/Motor Lighting Ballast Duplex Receptacle Total
SG403ATX 1625 RPM, 3.2 amps 0.8 amps 5.0 amps 9.0 amps
SG603ATX 1625 RPM, 6.6 amps 0.8 amps 5.0 amps 12.4 amps


SterilchemGARD® III Advance°
Model Number: SG403ATX, 4-foot
SG603ATX, 6-foot

NSF Classification: Class II Type B2
Cabinet Type: Console


Site Preparation
Electrical System
SG403ATX and SG603ATX

  • 115V-1 Phase, 60 Hz.
  • One 14-foot power cord with 20-amp plug, NEMA 5-20P.
  • Unit is UL listed for electrical, fire and personal safety.
  • Two ground fault interrupt protected interior duplex receptacles at 5.0 total amps, controlled by a circuit breaker switch.

Filtration System

  • Exhaust and supply filters are standard HEPA filters with an efficiency of 99.99% when filtering particles of 0.3 micrometers.


Exhaust Transition (ET)
For Hard Connection

The Exhaust Transition (ET) provides a hard connection, or airtight mating of the cabinet to the exhaust duct. The ET403-TX and ET603-TX are designed for the 4’ and 6’ cabinets, respectively. The Exhaust Transition incorporates the following features:

  • Rectangular base is easily installed on top of the cabinet over the exhaust filter, requiring only six fasteners to provide an airtight seal.
  • Design enables accurate filter leak checking through a convenient access panel on the sloping front surface.
  • Standard exhaust connection collar is 10" diameter, with optional 12" diameter available.

Exhaust Requirements
Cabinet
Models
Duct
Diameter
Sash
Height
Exhaust
Flow (CFM)
Pressure Requirement (wc)*
Clean 50% Loaded
SG403ATX 10" 8" 750-800 1.4" 1.8"**
12" 1.2" 1.6"
SG603ATX 10" 8" 1150-1200 2.0" 2.5"
12" 1.5" 2.0"
*Water column (wc) suction is measured directly above the exhaust transition before any dampers, elbows or other restrictions.
** Note: With airtight damper assembly add 0.1" (wc) to the above static pressure requiremants.


SterilchemGARD® III Advance°
Hard Connection Accessories

1. Low profile, slanted front exhaust transition with filter access panel and standard 10" (12" optional) round exhaust collar.

2. Optional airtight damper seals the cabinet during decontamination and increases the overall height by 73/4".

 

Warranty
The Baker Company, Inc., expressly represents and warrants all goods (a) to be as specified (and described) in The Baker Company catalogs and literature, and (b) to be free under normal use, service and testing (all as described in The Baker Company catalogs and literature) from defects in material and workmanship for a period of thirty-six months from the invoice date. The exclusive remedy for any breach or violation of this warranty is as follows: The Baker Company, Inc., will F.O.B. Sanford, Maine, furnish without charge repairs to or replacement of the parts or equipment which proved defective in material or workmanship. No claim may be made for any incidental or consequential damages.

THIS WARRANTY IS EXPRESSLY IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE UNLESS OTHERWISE AGREED IN WRITING SIGNED BY THE BAKER COMPANY. (THE BAKER COMPANY SHALL NOT BE RESPONSIBLE FOR ANY IMPROPER USE, INSTALLATION, SERVICE OR TESTING OF THE GOODS.)

Purchase Specification
SterilchemGARD® III Advance°
Class II, Type B2
Biological Safety Cabinet

1. Manufacturer shall provide a certified copy of the personnel, product and cross-contamination (biological) tests, equivalent to or more demanding than as specified in NSF International Standard #49, performed on the unit selected from the corresponding statistical sample. Tests may be witnessed by a representative of the purchaser.
2. Cabinet shall have Momentum Air Curtain downflow velocity profile, i.e., a higher velocity of downflow behind the viewscreen relative to downflow velocity over the work surface.
3. High velocity return air slots shall be located at each end of the front access opening. These slots help to prevent air from being drawn into the work area along the edges of the side wall and from escaping the work area to the ambient environment.
4. The sliding viewscreen shall be slanted at an angle of 10º from vertical, capable of moving to a fully closed position during shutdown periods.
5. Viewscreen shall be constructed of 1/4" safety plate glass, with a maximum opening of 18" for equipment loading.
6. All biologically contaminated ducts, plenums and work area side walls shall be permanent metal construction and maintained under negative pressure or enclosed within a negative pressure zone.
7. Cabinet shall be capable of automatically handling a 60% minimum increase in filter loading without reducing total air delivery by more than 10%. With the use of the speed controller, the cabinet shall be capable of handling a 180% increase in filter loading. Test data to verify these capabilities shall be available upon request.
8. Interior work area shall be 271/4" high.
9. Cabinet shall have UniPressure™ PreFlow System, designed to provide more uniform airflow to the supply filter.
10. Supply and exhaust HEPA filters shall be front-loading.
11. A plenum assembly shall be provided to allow the supply filter to be directly clamped to the plenum against a closed-cell neoprene gasket. Plenum applies force to full perimeter of filter, rather than point force.
12. Complete unit shall be listed by Underwriters Laboratory (UL) for electrical, fire and personal safety.
13. Sliding sash high velocity air return slots shall be provided along the entire top edge of the work area to prevent migration of contamination behind the sliding glass zone.
14. Audible and visual alarm system shall be provided to indicate low exhaust airflow, and shall be interlocked with unit supply blower to prevent pressurization of the work zone.
15. Cabinet shall have a microprocessor-based membrane control panel mounted on the front of the cabinet facing down toward the user when sitting at the unit.
16. Unit shall have an audible alarm and a flashing LED to indicate when the sliding viewscreen is in an unsafe position. An alarm mute switch shall be provided on the front-mounted cabinet control panel to allow the operator to mute the alarm tone for brief adjustments. The alarm shall automatically reset after five minutes if the viewscreen remains in an unsafe position.
17. Intake velocity through the 8" front access opening shall be minimum of 100 FPM.
18. Cabinet exterior construction: seal panels of 16-gauge cold-rolled steel and dress panels of 18-gauge cold-rolled steel, baked enamel finish Perma White™.
19. Cabinet interior (work area) construction: one-piece 16 gauge, Type 304 stainless steel, with a smooth, 7/16" radius between rear and side walls, and easily cleanable, radiused corners on the work surface tray.
20. Work area side walls and rear wall to be one-piece construction. A straight back wall shall be provided to maximize work area and easily accommodate laboratory equipment.
21. Work area shall be provided with two GFI protected duplex outlets with drip-proof covers and circuit breakers.
22. Lower front work area airfoil shall be provided to improve access opening containment capability.
23. Cabinet shall have a unitized drain pan with 7/16" radius on all sides and a fully-removable work surface and work surface supports to facilitate cleaning.
24. Stainless steel air diffuser filter protector provided in work area.
25. Unit must be listed by NSF International as meeting Standard #49.
26. Each unit, before shipping, shall have a complete physical test to assure cabinet meets Class II requirements. A copy of this test will be provided with the unit.
27. Speed controller shall automatically compensate for line voltage changes to maintain constant voltage to motor while allowing for manual adjustments to accommodate filter loading.
28. A single power cord and plug shall be provided for electrical power source.
29. One petcock shall be provided on the right wall inside the cabinet. Three additional capped penetrations shall be provided for addition of future petcocks, two on the left wall and one on the right wall.
30. All external plumbing connections to the petcocks and the drain pan shall be made standard through the bottom of the cabinet and not the sides, allowing zero clearance between the unit and the building walls or equipment to its right and left. Factory option for plumbing through the back or top of cabinet.
31. Unit shall have standard HEPA filters with an efficiency of 99.99% when filtering particles of 0.3-micron size.
32. Unit shall have a washable, re-useable, intake prefilter.
33. Unit shall have (optional) ultraviolet light with a shutoff safety feature when the viewscreen is raised.
34. Four-foot unit shall exhaust no more than 800 CFM of air with an average 105 FPM calculated air intake through front access opening at 8" operating level. Six-foot unit shall exhaust no more than 1200 CFM of air with an average 105 FPM calculated air intake through front access opening at 8" operating level.
35. Unit shall have electronic ballasts for UV and fluorescent lighting to provide longer life and lower heat output.
36. Unit shall have a telescoping stand that allows the work surface height to be set at either 30" or 36".
37. Unit shall have a removable armrest across the cabinet front.

 